- In general please be careful to use the word "efficiency" because this is not actual efficiency of the INTT but just a feasibility test of the efficiency estimation by using only 2 layers.
- Slide 8, 9: Approved. Add titles and units (cm) for X, Y axis. Need to explain this is single muon MC in the legend. Also need to explain what is the difference of the point color means. Don't change the color of the vertex point.
- Slide 10: Approved. Don't use the work "efficiency". "Num of good pairs/Num of outer clusters" would be better? (Cheng-Wei will give you a better wording.) Need the unit of Residual (cm). Change the unit of window from mm to cm to be same as Residual unit. Need more information in the legend (single Muon MC, z-vertex, Incident direction etc.)
- Slide 11: Approved. Modify as the same way as suggested for the plot in slide 10.
- Slide 13: Approved. Modify as the same way as suggested for the plot in slide 10.
Comment: How you can explain the 10% reduction of efficiency?
- Slide 14: Approved. Set Y range from 0 to 800. Remove the current legend and put more informative legend (explanation of black and red histograms, beam, z-vertex etc.) Don't use "vec_outer.Phi()", but use "Phi". Also need the unit of phi.
Comment: It would be great if you combine two plots into one plot (The ratio at the bottom), but if it's difficult, it would be okay to have two separate plots.
- Slide 15: Approved. Modify as the same say as the plot slide 14. Y axis range of the left panel should be 0 to 3000.
- Slide 19, 20: These event display plots will be removed from the WIP request.
- Slide 21 is not for WIP request, but just an internal plot requested by Yasuyuki.
- Slide 22, 23: Approved. Modify as the same way as suggested for the plot slide 10. The Y axis rage of the right plot should be same for slide 10, 11, 13, 22, 23. Need to set a proper Y axis range for the left plot of slide 23.
- Slide 24: Approved. Window need the unit and it should be cm unit. Need to emphasize "This is still work in progress" in the plot.